RecallDepth

GENERAL LAPAROSCOPIC PACK-LF , Model No UTGL53S-06 UTGL53S-07 UTGL53S-08

Company
American Contract Systems Inc
Recall Initiated
February 6, 2025
Posted
March 14, 2025
Recall Number
Z-1357-2025
Quantity
534 kits
Firm Location
Tiffin, OH

Reason for Recall

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Distribution

US Nationwide distribution in the state of OH.

Lot / Code Info

UDI-DI 00191072147196 Lots 941241 899241 70-050873 70-051230 70-051578 70-052009A 70-052884

Root Cause

Process control

Action Taken

The affected distributor was notified via email on 2/16/25, and an updated letter was provided on 2/18/25. The distributor was instructed to notify their end users. Customers were provided with a list of impacted components which are included in the affected trays. Customers were provided with a labeling template and instructed to add warning labels to affected procedure trays to ensure the affected components are discarded.

More recalls by American Contract Systems Inc

View all recalls by this company →