GENERAL ANGIO PACK , Model Nos LLAP75G LLAP75G-01 LLAP75G-02 LLAP75G-03
- Company
- American Contract Systems Inc
- Recall Initiated
- February 6, 2025
- Posted
- March 14, 2025
- Recall Number
- Z-1340-2025
- Quantity
- 956 kits
- Firm Location
- Tiffin, OH
Reason for Recall
ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.
Distribution
US Nationwide distribution in the state of OH.
Lot / Code Info
UDI-DI 00191072219510 Lots 70-050850 70-051285 70-051779 70-051779A 70-052186 70-052669 70-053022
Root Cause
Process control
Action Taken
The affected distributor was notified via email on 2/16/25, and an updated letter was provided on 2/18/25. The distributor was instructed to notify their end users. Customers were provided with a list of impacted components which are included in the affected trays. Customers were provided with a labeling template and instructed to add warning labels to affected procedure trays to ensure the affected components are discarded.
More recalls by American Contract Systems Inc
AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as:
1) A-C SUPPLEMENT PACK -...
Aug 28, 2025
AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as:
1) ANTERIOR HIP PACK, MODE...
Aug 28, 2025
AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as:
1) ROBOT HYST PACK, MODEL/...
Aug 28, 2025