3M Attest Super Rapid Readout Biological Indicator, REF 1492V; used to qualify or monitor dynamic-air-removal steam sterilization cycles
- Recall Initiated
- August 22, 2023
- Posted
- September 26, 2023
- Recall Number
- Z-2637-2023
- Quantity
- 79500 devices
- Firm Location
- Saint Paul, MN
Reason for Recall
Specific lots of 3M Attest Super Rapid Readout Biological Indicators (1492V), 3M Attest Super Rapid 5 Steam-Plus Challenge (41482V, 41482VF), and 3M Attest Super Rapid Readout Steam Challenge (1496V) may contain biological indicators with a cap that melts, deforms and/or cracks following a steam sterilization cycle of 132C and 135C.
Distribution
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Belgium, Canada, China, Colombia, Hong Kong, Italy, Japan, Peru, Poland, South Korea, Spain, Taiwan.
Lot / Code Info
UDI/DI: 50707387762969 (shipper), 30707387762965 (carton); Lot codes: 33P3X3, exp. 12/29/2024; 33P4NH, exp. 01/01/2025
Root Cause
Process control
Action Taken
3M issued an Urgent product recall notice to its consignees on 08/24/2023 by email and US mail. The notice explained the problem with the product and requested the following: Discontinue distribution of the product and dispose of the product. Distributors were directed to notify their customers. The firm is offering replacement with the return of the completed response form.