RecallDepth

Various Pacemaker Packs, Cardio Thoracic Packs

Company
American Contract Systems, Inc.
Recall Initiated
June 16, 2022
Recall Number
Z-1691-2022
Quantity
24 units
Firm Location
Kansas City, MO

Reason for Recall

Product was exposed to multiple sterilization cycles without validation for multiple exposures.

Distribution

Distribution within the United States to health systems and distributors in AZ, FL, IA, IL, MA, MN, MO, OH, NE, PA, RI, SD, and TX

Lot / Code Info

Tray Number, Sterilization Lot, Lot Number, Bag Serial Number 1) LSREPCUSTOML, 2106181, 835211, 44251067 2) LSRCVBASING, 2107102, 817211, 64116041 3) LSRSMVEINB, 2107221, 810211, 64119459 4) LSRCFHENDAE, 2107171, 807211, 64116602 5) UICT16E, 2108183, 771211, 15066268 6) BHOH94I, 2104141, 897211, 63987116-006 7) BHOH95, 2104301, 902211, 44123693-005 8) ANCV60AI, 2105201, 860211, 63956881-006 9) UDIP97AD, 2105201, 860211, 14858003-004 10) CEFS17I, 2105262, 855211, 44132805-005 11) FHPM10P, 2106092, 840211, 44129541-005 12) SLCV01E, 2107081, 812211, 64110741-006 13) SLEP03F, 2108271, 761211, 44314505-005 14) SLEP03F, 2108271, 761211, 44314504-005 15) CXCV91X, 2108311, 757211, 44315778-005 16) MECV98V, 2108312, 757211, 44313827-005 17) SAPA90X, 2108311, 758211, 64204754-006 18) SLCV01E, 2109161, 742211, 64208360-006 19) MEEP10F, 2109202, 740211, 64220361-006 20) UMOH42C, 2105072, 876211, 64061399-006 21) BHON40H, 2108122, 778211, 64095875-006 22) BHPM60F, 2108122, 779211, 44216056-005 23) BHCA27C, 2109072, 754211, 63997844-006 24) BHPM60F, 2109292, 728211, 44353475-005

Root Cause

Process change control

Action Taken

The recalling firm began notifying customers of the recall beginning June 13, 2022. Customer notification was initially made via email and an "Urgent Medical Device Recall" letter dated June 16, 2022. Customers are asked to review inventory and segregate affected product, which should be identified with the Bag Serial Number. Affected product should be destroyed, with evidence of destruction provided back to the recalling firm. A Field Action Response Form should be completed and returned to angela.bunn@owens-minor.com as soon as possible, even if the customer has no inventory. If you have any questions, please call 816-920-5846.

More recalls by American Contract Systems, Inc.

View all recalls by this company →