OAKWORKS Inc.Lithotripsy/Urology Table: CFLU401, Catalog Number: 75216-T01 - Product Usage: Oakworks Imaging Tables are radiographic tables intended for use with mobile or compact stationary C-arm Imaging Systems. It is ideally suited for pain management imaging and therapeutic procedures. These tables are suitable to use for diagnostic x-ray imaging and imaging during therapeutic procedures such as spinal injections, vertebroplasty procedures and other pain management procedures. The CFLU Table is designed for Lithotripsy and Urology procedures. It is intended to be operated by a healthcare professional in a medical environment. No special training is required but a review of the Safety Instructions is important for the safety of the operator and patient.
This recall has been terminated (originally issued May 2, 2019).
- Company
- Oakworks Inc
- Recall Initiated
- May 2, 2019
- Terminated
- July 8, 2019
- Recall Number
- Z-1411-2019
- Quantity
- 3 units
- Firm Location
- New Freedom, PA
- Official Source
- View on FDA website ↗
Reason for Recall
Table assembly error with incorrect actuators for the lateral and longitudinal table movements
Distribution
US Nationwide distribution in the state of SC.
Lot / Code Info
Serial numbers:FLL745088, FLL745089, and FLL745090.
Root Cause
Nonconforming Material/Component
Action Taken
Oakworks issued a Medical Device Advisory Letter on 5/02/19 advising of the problem, and action to take: Immediately examine your inventory of OAKWORKS Inc., Multi-Movement Imaging Table: CFLU401 to identify and locate all affected product(s). 2) Do not distribute these affected product(s). Please quarantine the table(s). If further distribution of this product, contact your customer(s) with a copy of this letter and all applicable response forms. A service visit will replace the incorrect actuators on the table(s). Questions or concerns, please contact our Customer Service Department at 1-800-558-8850 or 1 717-235-6807