OAKWORKS Inc.Lithotripsy/Urology Table: CFLU401, Catalog Number: 75216-T01 - Product Usage: Oakworks Imaging Tables are radiographic tables intended for use with mobile or compact stationary C-arm Imaging Systems. It is ideally suited for pain management imaging and therapeutic procedures. These tables are suitable to use for diagnostic x-ray imaging and imaging during therapeutic procedures such as spinal injections, vertebroplasty procedures and other pain management procedures. The CFLU Table is designed for Lithotripsy and Urology procedures. It is intended to be operated by a healthcare professional in a medical environment. No special training is required but a review of the Safety Instructions is important for the safety of the operator and patient.

This recall has been terminated (originally issued May 2, 2019).

Company
Oakworks Inc
Recall Initiated
May 2, 2019
Terminated
July 8, 2019
Recall Number
Z-1411-2019
Quantity
3 units
Firm Location
New Freedom, PA
Official Source
View on FDA website ↗

Reason for Recall

Table assembly error with incorrect actuators for the lateral and longitudinal table movements

Distribution

US Nationwide distribution in the state of SC.

Lot / Code Info

Serial numbers:FLL745088, FLL745089, and FLL745090.

Root Cause

Nonconforming Material/Component

Action Taken

Oakworks issued a Medical Device Advisory Letter on 5/02/19 advising of the problem, and action to take: Immediately examine your inventory of OAKWORKS Inc., Multi-Movement Imaging Table: CFLU401 to identify and locate all affected product(s). 2) Do not distribute these affected product(s). Please quarantine the table(s). If further distribution of this product, contact your customer(s) with a copy of this letter and all applicable response forms. A service visit will replace the incorrect actuators on the table(s). Questions or concerns, please contact our Customer Service Department at 1-800-558-8850 or 1 717-235-6807

More recalls by Oakworks Inc

View all recalls by this company →