Ultrasound EA (Ergonomic Advantage) table versions; G3 Multi-Specialty Echocardiography and Multi-Specialty with Electric Access Back Supports [left & right}

This recall is currently active, issued January 13, 2024. It was issued by Oakworks Inc.

Company
Oakworks Inc
Recall Initiated
December 5, 2023
Posted
January 13, 2024
Recall Number
Z-0729-2024
Quantity
20 units
Firm Location
New Freedom, PA
Official Source
View on FDA website ↗

Reason for Recall

Backrest platform stress fracture - the table is not operational in an inclined position

Distribution

US Nationwide Distribution

Lot / Code Info

(UDI): 00817463025422 S/N: USM839457 USM839458 USM839459 USM839460 USM839461 USM818212 USM833897 USW813999 USM839614 USM839615 USM814871 USM833507 USM831797 USM823858 USM822620 USM824500 USM834529 USM788091 USM815794 USM832170

Root Cause

Device Design

Action Taken

Oakworks Inc. issued "URGENT: MEDICAL DEVICE RECALL/CORRECTION" letter on 12/5/23 via Certified Mail/Email to Distributors and End Users. Letter states reason for recall, health risk and action to take: 1. Immediately examine your inventory of OAKWORKS Inc. Ultrasound EA Table with Echocardiography, or Electric Access Back Supports [left & right] to identify and locate all affected product(s). 2. For tables in use, they remain fully functional if the backrest platform stress fracture has not occurred. For tables that have the stress fracture or break, only use in a flat position until the backrest is replaced. 3. Implement the field action as described below for the product listed in the Product and Distribution Information Table. 4. Enclosed is a Medical Device Advisory Notice Response Form for each product we believe you have. Please complete this form in its entirety and return to OAKWORKS Inc., within 48 hours of receipt of this advisory notice via any of the options listed below. MAIL: Oakworks Inc. EMAIL: customerservice@oakworks.com Attn: Customer Service MDAN Response 923 East Wellspring Rd. New Freedom, PA 17349 Action by OAKWORKS Inc.: Upon receipt of the Medical Device Advisory Notice Response Form(s), and to minimize the downtime of your OAKWORKS Ultrasound EA table, we will setup a service visit to replace the backrest (if a stress fracture has occurred) or install a reinforcement plate (if a stress fracture has not occurred) on the table(s). The correction installation takes approximately 30 minutes. The replaced backrest will be disposed of by the service provider. Additional Information: If you have any questions or concerns, please contact our Customer Service Department at 800-558-8850 or 717-235-6807 Monday through Friday, 8:00 AM to 5:00 PM, Eastern Time.