cobas 8000 Modular Analyzer Series; Part Numbers: 1. 05964067001 cobas 8000 c 502 module

This recall has been terminated (originally issued July 18, 2017).

Company
Roche Diagnostics Corporation
Recall Initiated
July 18, 2017
Terminated
July 13, 2018
Recall Number
Z-0910-2018
Quantity
245
Firm Location
Indianapolis, IN
Official Source
View on FDA website ↗

Reason for Recall

A warning is being added to the cobas 6000 analyzer and the cobas 8000 modular analyzer series Operators Manual to prevent operator injury during maintenance.

Distribution

Distributed in all 50 states: AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MN, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, the District of Columbia, and Puerto Rico.

Lot / Code Info

Serial Numbers: 16C4-09 1475-01 1476-05 15A1-03 15A1-05 16B3-10 1474-06 1138-10 1140-10 1485-06 1590-01 1475-05 1484-05 1484-06 1012-06 1482-02 1481-10 1254-07 1487-07 1482-09 1482-10 16B7-08 16B8-06 1010-03 1020-09 15A2-02 1247-02 1267-06 15A2-03 1267-01 1369-02 16B5-08 16B3-03 15A5-04 15A5-03 1124-03 1125-04 1591-04 1591-05 1474-09 1474-08 1009-05 1021-08 1139-03 1140-09 1267-02 1267-03 1140-06 16B1-02 16B1-03 1138-04 1138-09 16B4-03 16C3-03 1485-07 1597-09 1011-10 1018-10 16C7-08 16C4-07 1475-02 1475-04 1476-07 16C4-10 16C5-10 15A4-04 15A4-05 1267-07 1474-07 1266-08 1267-09 1256-02 1256-03 1125-05 1369-08 15A6-04 15A7-03 1020-05 1124-04 1123-08 1485-09 1597-04 15A0-03 16C1-05 16C1-06 16C1-07 1267-08 1594-02 1594-04 1594-05 1594-06 1595-01 1595-06 1597-02 1597-03 15A3-01 15A4-01 16B0-02 16B0-03 16B0-04 16B8-03 16B8-04 16B5-09 16B8-01 16B2-07 16B2-08 1266-06 1266-07 1591-01 1591-02 16B1-07 16B2-05 16B2-06 16B1-01 16B4-10 16B5-01 15A6-02 15A6-03 15A6-06 1253-06 16B5-04 16B5-05 1592-01 1592-06 1592-07 16C7-09 16D0-10 1138-07 1139-05 16B1-04 16B1-05 1369-04 1124-05 1124-06 1139-01 1139-02 1253-04 1253-03 1251-03 1251-07 1482-03 1483-07 16C2-03 16B0-05 16B0-06 16B0-01 1124-09 1124-10 15A5-02 16A9-06 1017-04 1122-04 16C5-03 1483-09 1483-10 15A4-06 1012-08 1131-02 1131-03 1137-08 1139-04 1592-10 1597-06 1597-07 1597-08 16B4-01 16B4-02 16B6-05 16C3-04 16C4-06 16B7-05 16B7-06 1476-04 1475-03 16B8-05 16B8-09 1139-06 1591-07 16C1-04 1135-02 1137-07 1266-09 1266-03 1267-04 1472-05 1592-09 1592-08 1138-05 1266-05 1254-09 1254-10 1485-05 1019-03 1132-03 1472-02 1474-10 1132-04 1132-05 1021-04 1021-06 1247-01 1253-07 1472-03 1472-04 1477-06 1482-06 16C8-02 16C8-03 16B1-06 16B7-07 16B5-07 16B8-07 15A1-06 15A1-08 15A5-01 16B2-09 16B3-02 15A6-05 16A9-05 1123-05 1125-03 1251-06 1266-10 1140-07 1140-08 1254-08 1125-02 1132-09 16C1-10 16C1-09 1267-05 1253-05 1483-06 1483-08 1132-07 1133-09 16C1-02 1484-10 1484-09 1484-08 16C5-01 1015-07 1018-09 1123-06 16B6-04

Root Cause

Device Design

Action Taken

Actions Taken by Roche Diagnostics: A warning will be added to the cobas 6000 analyzer series Operators Manual and the cobas 8000 modular analyzer series Operators Manual. Actions to be Taken by the Operator: Operators are advised not to overtighten the thumb screws when replacing the USM Cover 3 after maintenance. When removing the screws for maintenance, avoid touching the edge of the USM Cover 3. Enclosure: Fax back form 7727-00-0717 Actions Required: " Do not overtighten the thumb screws when replacing the USM Cover 3 after maintenance. When removing the screws for maintenance, avoid touching the edge of the USM Cover 3. " Complete the attached fax form and fax it to 1-888-276-5901. " File or post this Urgent Medical Device Correction (UMDC) in an appropriate location within your laboratory, for future reference.