RecallDepth

D-RAD Self-Tapping Locking Screw intended for the fixation of fractures involving the distal radius. smith&nephew D-RAD 2.4MM X 10MM S-T LOCKING SCREW, REF 74692410, QTY (1), STERILE R

This recall has been terminated (originally issued March 20, 2018).

Company
Smith & Nephew Inc
Recall Initiated
March 7, 2018
Posted
March 20, 2018
Terminated
October 31, 2018
Recall Number
Z-1317-2018
Quantity
59 units
Firm Location
Memphis, TN
Official Source
View on FDA website ↗

Reason for Recall

One lot of D-RAD Self-Taping Locking Screws used with the Distal Radius Fracture Kit were machined out of specification. Screws measured over tolerance within the head thread form by 0.002-0.003 inch.

Distribution

US, Puerto Rico, Malaysia

Lot / Code Info

Lot Number 17GM04522

Root Cause

Under Investigation by firm

Action Taken

The firm initiated their recall by letter and email on 03/07/2018. The firm requested return of the product.