RecallDepth

Siemens RAPIDPoint 500 v2.2.2A Software Upgrade Kit; SMN 11066719, Software Version 2.2A The RAPIDPoint 500 system is designed for professional use in a point-of care or laboratory environment. This system tests blood gases, electrolytes, metabolites, total hemoglobin, and hemoglobin derivatives in arterial, venous, and capillary whole blood samples. The following parameters are tested: pH, pC02, p02, Na+, K+, Ca++, Cl-, glucose, lactate, tHb, F02Hb, FCOHb, FMetHb, FHHb, nBili

This recall has been terminated (originally issued August 19, 2016).

Company
Siemens Healthcare Diagnostics Inc
Recall Initiated
July 5, 2016
Posted
August 19, 2016
Terminated
August 2, 2017
Recall Number
Z-2601-2016
Quantity
76
Firm Location
Norwood, MA
Official Source
View on FDA website ↗

Reason for Recall

Some v2.2.2 upgrade kits include a dialysate mode which not cleared/approved for shipment in the United States

Distribution

Nationwide Distribution to AL, CA, MA, NJ, OH, TX, WA, PA

Lot / Code Info

Software Version 2.2A

Root Cause

Under Investigation by firm

Action Taken

Siemens sent a Customer Notification dated July 2016 to all affected domestic (US) customers via Federal Express on July 5, 2016. Customers were asked to take the following actions: Please do not use the dialysate mode. If you have enabled this feature, please disable it by following these instructions: 1. At the System screen, select Setup. 2. Select Secured Options > Analysis Options. 3. Unselect Dialysate. 4. Select the Continue button 3 times. Please review this letter with your Medical Director. Complete and return the Field Correction Effectiveness Check attached to this letter within 7 days. Customers with questions were instructed to contact their Siemens Customer Care Center or their local Siemens Technical Support representative. For questions regarding this recall call 781-269-3000.

More recalls by Siemens Healthcare Diagnostics Inc

View all recalls by this company →