Spiros CLOSED MALE LUER w/RED CAP, 25 UNITS, REF: 011-CH2000SC-25; 76 cm (30") Appx 3.2 ml, 20 Drop Admin Set w/Integrated ChemoClave Drip Chamber, Spiros w/Purple Cap, Bag Hanger, REF: 011-CH3185; 46 cm (18") Appx 7.4 ml, Trifuse Add-On Set w/2 Spiros w/Purple Caps, Bag Spike, 3 Clamps (Blue, 2 Red), Dry Spike Adapter, REF: 011-CH3266; 76 cm (30") Appx 3.3 ml, Admin Set, 2 Spiros w/Purple Cap, 20 Drop In-Line Drip Chamber, Bag Hanger, REF: 011-CH3446; 61 cm (24") Appx 8.2 ml, Bifuse Add-On Set w/Bag Spike, Spiros w/Purple Cap, 3 Clamps (1 Blue, 2 White), Check Valve, Dry Spike Adapter, REF: 011-CH3729; 25 UNITS of Spiros CLOSED MALE LUER, REF: 011-H2457; Appx 1.0 ml, Adattatore per Pompa con Spiros, REF: 011-H2629; 179 cm (70") Appx 11.5 ml, PUR Ambr. Deflussore con Spiros, Camera, BCV, Y-Clave e LL Gir., REF: 011-H2776; 180 cm (71IN)APPX 11.5ml,PUR AMBR,Spiros,CAMERA,REG.FLUSSO,BCV,Y-Clave, REF: 011-H2864-HSR; 181cm(71 IN)APPX11.6ml,PUR DEFLUSSORE,Spiros,20 DROP IN LINE,BCV,Clave, REF: 011-H2904; 38 cm (15 IN) APPX 2.9 ml, TRIFUSE EXT SET, 3 MicroClave, Spiros, PURPLE CAP, REF: 011-H3449; 42cm(16.5 IN)APPX 2.6ml,PUR TRANSFER SET,Clave ADDITIVE PORT,BCV,Spiros, REF: 011-H3597; 41cm(16 IN)APPX 2.9ml,PUR SET PERFORATORE,Clave,VALVOLA,Spiros, REF: 011-H3628; 41cm(16 IN)APPX 3.1ml,SET AMBRATO,PERFORATORE,Clave,VALVOLA,Spiros, REF: 011-H3629; 42cm(16.5 IN)APPX 5.6ml,SET,PERFORATORE,Clave,FILTRO,VALVOLA,Spiros, REF: 011-H3630; 109 cm(43 IN) APPX 7.2 ml,KIT,EXT SET W/Spiros,STOPCOCK,5 Clave,4-GANG STOPCOCK, REF: 011-MC330632; 2 Ext Sets w/5-Gang Stopcock, 9 Clave, Spiros, Stopcock, REF: 011-MC33242; Kit for multiple transfers, REF: 011-N1000; Kit per il riempimento pompe elastomeriche con SPIROS / Kit for elastomeric pump priming with SPIROS, REF: 011-N1000-P2; APPX 2.3ml, ZYTO-VERBINDUNGSLEITUNG (ZYTO-CONNECTING LINE), REF: 011-N1664SP; Appx 1.0 ml, Adaptador universal para sistemas IV con Spiros, REF: 034-H2629; Spinning Spiros Closed Male Luer, Red Cap (No Flange), REF: CH2000S-NR; 15" (38 cm) Appx 5.2 ml, Bifuse Add-On Set w/Bag Spike, Spiros, 3 Clamps (Blue, 2 White), Dry Spike Adapter, REF: CH3187; 14" (36 cm) Appx 8.0 ml, Quinfuse Add-On Set w/4 Spiros, Bag Spike, 5 Clamps (Blue, 4 Red), Dry Spike Adapter, REF: CH3225; 20" (51 cm) Appx 9.5 ml, Quadfuse Add-On Set w/3 Spiros, Bag Spike, 4 Clamps (Blue, 3 Red), Dry Spike Adapter, REF: CH3226; 16" (41 cm) Appx 6.3 ml, Trifuse Add-On Set w/2 Spiros, Bag Spike, 3 Clamps (Blue, 2 Red), Dry Spike Adapter, REF: CH3227; 30" (76 cm) Appx 4.1 ml, Yellow 20 Drop Admin Set w/15 Micron Filter, Spiros, REF: CH3235; 28" (71 cm) Appx 5.1 ml, Add On Set w/Clave Spike, Spiros, Clamp, REF: CH3354; 16cm(6 IN)APPX 2.4ml,SET,MicroClave Clear,STOPCOCK,DROP-IN ADD-ON,Spiros, REF CH3411: APPX 1.9 ml,13 cm (5 IN),BAG SPIKE,Spiros,ACCESS DEVICE w/Clave, REF: CH3502; 60 (152 cm) Appx 1.9 ml, Smallbore Ext Set w/MicroClave Clear, 0.2 Micron Low Protein Binding Filter, Spiros w/Red Cap, Clamp, REF: CH3603; Spinning Spiros Closed Male Luer, REF: K7040-001; APPX 0.76ml, 4-WAY STOPCOCK w/2 Clave, VIAL ADAPTER, 3 Spiros, REF: K7052-001; ZYTOmed Verab.-Set Onko II lichtschutz, REF: SN1205AG; ZYTOmed - Verabreichungs-Set Onko III lichtgeschtzt, REF: SN1206A; Conector macho cerrado Spinning Spiros, tapon purpura, REF: 034-CH2000S-PC
- Company
- Icu Medical, Inc.
- Recall Initiated
- August 31, 2020
- Recall Number
- Z-0706-2021
- Quantity
- 217522
- Firm Location
- Amanecer San Clemente, CA
Reason for Recall
Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patient blood, or fluid path air-in-line.
Distribution
Worldwide distribution - U.S. Nationwide distribution including in the states of FL, MD, LA, PA, IN, TX, NY, MO, VA, OH, GA, NC, IL, CA, AZ, AK, NE, DE, WI, CO, UT, ND, MS, MA, MN, WA, SC, OR, KS, MI, NH, NJ, KY, TN, ID, AR, CT, OK, AL, SD, ME, IA, VT, WY, DC, NV, RI, MT, NM, HI, WV and the countries of Italy, Belgium, United Kingdom, Canada, Germany, Argentina, Australia, Poland, Greece, Switzerland, Israel, France, Finland, Spain, Malaysia, Taiwan, Ireland, Luxembourg, Hong Kong , Brazil, Netherlands, Austria, Sweden, Turkey, Singapore, Slovakia, Croatia, South Africa, Romania, Thailand.
Lot / Code Info
Lots: 4630846, 4848415, 4734764, 4734940, 4752044, 4742983, 4764957, 4878172, 4808952, 4867502, 4804800, 4854732, 4783273, 4773067, 4807223, 4800168, 4877026, 4752842, 4773432, 4764194, 4727034, 4847780, 4876626, 4548846, 4573171, 4867154, 4515064, 4574113, 4743010, 4773247, 4848995, 4740817, 4774162, 4877106, 4902461, 4711498, 4719930, 4740835, 4798726, 4866446, 4894490, 4727444, 4741918, 4778013, 4800251, 4806881, 4743840, 4734281, 4775528, 4799761, 4763261, 4774080, 4790739, 4749875, 4794599, 4802672, 4859281, 4718351, 4727042, 4741435, 4774258, 4866217, 4875126, 4899256, 4774269, 4777289, 4849437, 4853031, 4847272, 4882644 ,
Root Cause
Process control
Action Taken
On 08/31/2020, Medical Device Recall notices were mailed to customers informing them to take the following actions: 1) Please discontinue the use and distribution of the affected product immediately. Check your inventory and quarantine all affected product at your facility. 2) Inform potential users of the product in your organization of this notification and complete the attached response form. Return the completed response form to the fax number or e-mail address on the form, even if you do not have the affected product. 3) ICU Medical has some lots of unaffected product available today and is actively increasing the amount of available inventory. In the event specific product is unavailable, consider use of the non-spinning Spiros or the ChemoLock CSTD as alternatives. Please contact customer service for product availability. 4) Return affected product using the return label provided with this letter. Contact Stericycle at 1-888-871-7108 (MF, 8am-5pm ET) if you have not received a return label or require additional labels for returning the affected product. 5) If you have distributed the product further, immediately notify your accounts that received the product identified in the Affected Product / Table 1 sections of this notification and ask them to contact Stericycle to obtain a response form. For further inquiries, please contact: Global Complaint Management, 1-844-654-7780 or ProductComplaintsPP@icumed.com; Customer Service, 1-866-829-9025, option 8 or, customerservice@icumed.com, (M-F, 8am-6pm CT) On 10/07/2020, the Medical Device Recall notice, with an expanded affected device scope, was sent to customers.