3.5mm Aresenal Screw Drill Bit , Ref: 330-35-001, Qty. 70, Rx Only, Non-Sterile, UDI: (01)00812926029696 - Product Usage: A drill bit facilitates implant insertion

This recall has been terminated (originally issued October 10, 2019).

Company
Trilliant Surgical, Llc
Recall Initiated
October 10, 2019
Terminated
April 27, 2022
Recall Number
Z-1892-2020
Quantity
5 drill bits
Firm Location
Houston, TX
Official Source
View on FDA website ↗

Reason for Recall

Labeling Error; specifically, drill bits from lots # TSL008495A are marked with the lot # of TSL008495. This could lead to parts being misidentified and impact device traceability.

Distribution

US Nationwide distribution in the state of CA. OUS: None

Lot / Code Info

Lot # TSL008495A

Root Cause

Labeling design

Action Taken

On 10/10/2019 the firm sent a "Urgent: Medical Device Recall" via: FedEx overnight to all consignees in receipt of the recalled product. In addition to informing consignees about the recall, they ask consignees to take the following actions: Check inventory stock for the identified device. 2. Locate and remove and/or discontinue use of identified device. 3. Return all products of the affected device to Trilliant Surgical in accordance with the included Return Response. 4. For questions regarding this communication, contact Trilliant Surgical at 800.495.2919. (Monday Friday, 8:00AM 5:00 CST)