8mm HTR Sterile Hammer Toe Reaming Kit, Material: Stainless Steel, Sterile: R, PX only, UDI: (01)00812926022345(17)231225 (10)TSL007202
This recall has been terminated (originally issued April 1, 2020).
- Company
- Trilliant Surgical, Llc
- Recall Initiated
- November 4, 2019
- Posted
- April 1, 2020
- Terminated
- April 28, 2021
- Recall Number
- Z-1635-2020
- Quantity
- 117
- Firm Location
- Houston, TX
- Official Source
- View on FDA website ↗
Reason for Recall
The firm has identified that parts from the kit, lot: TSL007202, have the potential to contain the incorrect contents of the 8mm HTR Sterile Hammer Toe Reaming Kit. The kit should contain (2) reamers total, (1) Concave Reamer and (1) Convex Reamer; there is the potential that kits from the identified lot may have (2) of the same reamer type as opposed to one (1) of each type.
Distribution
Distributors in 21 states: AR, AZ, CA, FL, GA, IA,IN, KS, KY, MN, MO, NC, ND, NJ, NY, OH, TN, TX, UT, WA, and WI.
Lot / Code Info
Part Number: 220-08-003, Lot Number: TSL007202;Expiration Date: 12/25/2023.
Root Cause
Mixed-up of materials/components
Action Taken
The recall strategy included sending complete recall communication and response forms as of 11/04/2019 via FedEx (Express Standard Overnight) to all consignees that showed possession / had been sold a device as of 10/30/2019. The recall communication included the following actions to be taken by the distributor/customer: 1. Check inventory stock for the identified device. 2. Open the shipper box* on the end opposite of the label and remove the internal packaging. *Note: Opening the shipper box does not disrupt the sterile barrier. The double sterile barrier is maintained within the internal packaging. 3. Through the transparent internal packaging, verify that the kit contains the appropriate quantity and type of each reamer; (1) Concave Reamer and (1) Convex Reamer. If product in possession IS identified as affected by containing (2) of the same reamer type, quarantine and/or discontinue use of the device and return the affected device(s) to the firm in accordance with the included Return Response (Acknowledgment and Receipt Form). If product in possession IS NOT affected (i.e. there is 1 Concave Reamer and 1 Convex Reamer within the packaging, complete and return the included Return Response (Acknowledgment and Receipt Form). THE INCLUDED RETURN RESPONSE IS REQUIRED EVEN IF AFFECTED PRODUCT IS NOT IDENTIFIED. For questions regarding the communication, the phone number is 800.495.2919.