RecallDepth

MEDLINE NON-STERILE PVP SOLUTION, packaged in kits as follows: OISC HAND BEAR, Model Number DYNJ902348L

This recall is currently active, issued July 12, 2023. It was issued by Medline Industries, Lp - Northfield.

Company
Medline Industries, Lp - Northfield
Recall Initiated
May 18, 2023
Posted
July 12, 2023
Recall Number
Z-2134-2023
Quantity
128 units
Firm Location
Northfield, IL
Official Source
View on FDA website ↗

Reason for Recall

Affected product lots were mislabeled as "STERILE PVP SLN" when they should have been titled "NON-STERILE PVP SOLUTION, STERILE PACKAGING".

Distribution

Worldwide distribution - US Nationwide and the country of Panama.

Lot / Code Info

UDI/DI (EA) (01)10195327286057, UDI/DI(CS) (01)40195327286058; Lot Numbers: 23EDA775, 23CDB962

Root Cause

Process control

Action Taken

Medline issued a recall notice on 06/02/2023 to its consignees via USPS first class mail. The notice explained the problem with the product and requested the destruction of the product. Medline identified additional products, lots, and consignees and notified their consignees by email on 06/09/2023. Upon receipt of the submitted response form each account will receive over-labels to affix to the affected kits via FedEx overnight, if applicable. distributors were directed to notify their customers and collect the returned product and return it to Medline.

More recalls by Medline Industries, Lp - Northfield