GEM FLOW COUPLER Monitor, PN 5156-00000-011

This recall is currently active, issued July 20, 2023. It was issued by Baxter Healthcare Corporation.

Company
Baxter Healthcare Corporation
Recall Initiated
June 19, 2023
Posted
July 20, 2023
Recall Number
Z-2208-2023
Quantity
122 units
Firm Location
Deerfield, IL
Official Source
View on FDA website ↗

Reason for Recall

An issue was identified where the LCD monitor display may not indicate the correct active channel. The issue occurs when the monitor is turned on using battery power, and channel B was selected as the last active channel prior to powering down.

Distribution

US

Lot / Code Info

GTIN 00844735006148, all serial numbers

Root Cause

Under Investigation by firm

Action Taken

Synovis Micro Companies Alliance (MCA), Inc., a subsidiary of Baxter International Inc., is issuing an Urgent Medical Device Correction to its consignees on 06/19/2023 by USPS First Class Mail. The notice explained the problem with the device, risk, and provided the following method to circumvent the problem: "The affected flow coupler monitors may continue to be used. To avoid encountering this issue, please use the monitor while connected to an AC power supply whenever possible. However, if using battery power, Baxter advises users to select channel A prior to shutting down the monitor. Upon powering the monitor back up, please check to ensure that the channel illuminated via the hardware button matches the LCD screen channel display. If a mismatch occurs, touch the channel display on the LCD screen or press the hardware button for channel B to resolve the issue. For general questions regarding this communication, please contact Synovis MCA at 800-510-3318 or 205-941-0111, between the hours of 8:00 am and 5:00 pm Central Time, Monday through Friday."

More recalls by Baxter Healthcare Corporation

View all recalls by this company →