MEDLINE UNITE FOOT & ANKLE Screw, Non-Locking, Mini T8 (2.0x8mm), REF MMSN2008
- Recall Initiated
- December 22, 2022
- Posted
- February 1, 2023
- Recall Number
- Z-1045-2023
- Quantity
- 420 units
- Firm Location
- Northfield, IL
Reason for Recall
The recall is due to observed intra-operative screw failure. When pre-drilled with 1.3mm drill bit for 2.0mm screws and 1.6mm drill bit for 2.4mm we have observed screw failure due to excessive torque.
Distribution
US Nationwide distribution.
Lot / Code Info
GTIN 10193489120387, Lot Number 810304
Root Cause
Device Design
Action Taken
Medline issued a MEDICAL DEVICE RECALL notice to its direct consignees on 01/05/2023 by US mail and its distributors on 01/07/2023 by email. The notice explained the problem and requested the following: "REQUIRED ACTION: 1. Immediately check your stock for the affected items. Quarantine all affected product. All consignment inventory will be addressed by your local Medline UNITE Sales Representative. 2. Please return the completed enclosed response form listing the quantity of affected product on hand. Even if you do not have any affected product please complete and return the form, as indicated on the response form. 3. If you have transferred possession of this product to another individual, to another department or another location in your organization you are required by FDA regulations to notify them of this recall communication. Have the customers return any affected product to you for collection and return to Medline Industries, LP. FURTHER ACTION: Medline is immediately recommending the use of our 1.6mm pre-drill for 2.0mm screws, and our 1.9mm pre-drill for the use of our 2.4mm screws and 2.7mm screws. If you have any questions, please contact 866-359-1704."