Medline medical procedure kits, labeled as follows: a) PARACERVICAL PUD TRAY, REF DYNDH1832; b) AHS PARACERVICAL/PUDENDAL KIT, REF DYNJRA2210; c) PARACERVICAL/PUDENDAL BLK KIT, REF DYNJRA9046; d) PARACERVICAL/PUDENDAL BLK KIT, REF DYNJRA9046H; Medical procedure convenience tray
This recall is currently active, issued May 29, 2024. It was issued by Medline Industries, Lp - Northfield.
- Recall Initiated
- March 1, 2024
- Posted
- May 29, 2024
- Recall Number
- Z-1918-2024
- Quantity
- 12994 units
- Firm Location
- Northfield, IL
- Official Source
- View on FDA website ↗
Reason for Recall
Trumpet Needle Guide ring can detach when excessive pressure is applied.
Distribution
US Nationwide distribution.
Lot / Code Info
a) DYNDH1832: UDI/DI 40195327330713 (case), 10195327330712 (unit), Lot Numbers: 23KBI682, 23KBT359, 23KBX629; b) DYNJRA2210: UDI/DI 40195327424047 (case), 10195327424046 (unit), Lot Numbers: 23JBX245, 23LBN930; c) DYNJRA9046: UDI/DI 40195327338610 (case), 10195327338619 (unit), Lot Numbers: 23JBX367, 23JBX368, 23JBX369, 23JBX370, 23LBA896, 23LBH751, 23LBL612, 23LBV325, 23LBV326; d) DYNJRA9046H: UDI/DI 40195327338610 (case), 10195327338619 (unit), Lot Numbers: 23JBX367, 23LBL612, 23LBV325;
Root Cause
Process control
Action Taken
Medline issued a correction notice to its consignees on 03/01/2024 via email and USPS first class mail. The notice explained the issue and requested the consignee identify and quarantine all affected products, and respond to the notice. Once the response was received Medline would provide labels to place over the affected product in stock. If the product was further distributed the party which distributed the product was directed to notify their customers.