RecallDepth

Medline medical procedure kits, labeled as follows: a) PARACERVICAL PUD TRAY, REF DYNDH1832; b) AHS PARACERVICAL/PUDENDAL KIT, REF DYNJRA2210; c) PARACERVICAL/PUDENDAL BLK KIT, REF DYNJRA9046; d) PARACERVICAL/PUDENDAL BLK KIT, REF DYNJRA9046H; Medical procedure convenience tray

This recall is currently active, issued May 29, 2024. It was issued by Medline Industries, Lp - Northfield.

Company
Medline Industries, Lp - Northfield
Recall Initiated
March 1, 2024
Posted
May 29, 2024
Recall Number
Z-1918-2024
Quantity
12994 units
Firm Location
Northfield, IL
Official Source
View on FDA website ↗

Reason for Recall

Trumpet Needle Guide ring can detach when excessive pressure is applied.

Distribution

US Nationwide distribution.

Lot / Code Info

a) DYNDH1832: UDI/DI 40195327330713 (case), 10195327330712 (unit), Lot Numbers: 23KBI682, 23KBT359, 23KBX629; b) DYNJRA2210: UDI/DI 40195327424047 (case), 10195327424046 (unit), Lot Numbers: 23JBX245, 23LBN930; c) DYNJRA9046: UDI/DI 40195327338610 (case), 10195327338619 (unit), Lot Numbers: 23JBX367, 23JBX368, 23JBX369, 23JBX370, 23LBA896, 23LBH751, 23LBL612, 23LBV325, 23LBV326; d) DYNJRA9046H: UDI/DI 40195327338610 (case), 10195327338619 (unit), Lot Numbers: 23JBX367, 23LBL612, 23LBV325;

Root Cause

Process control

Action Taken

Medline issued a correction notice to its consignees on 03/01/2024 via email and USPS first class mail. The notice explained the issue and requested the consignee identify and quarantine all affected products, and respond to the notice. Once the response was received Medline would provide labels to place over the affected product in stock. If the product was further distributed the party which distributed the product was directed to notify their customers.

More recalls by Medline Industries, Lp - Northfield

View all recalls by this company →