iTotal CR Knee Replacement System Model Number: TCR-111-1111-010102 - Product Usage: The iTotal CR Knee Replacement System is intended for use as a total knee replacement in patients with knee joint pain and disability whose conditions cannot be solely addressed by the use of a prosthetic device that treats only one or two of the three knee compartments, such as a unicondylar, patellofemoral or bicompartmental prosthesis.

Company
Conformis, Inc.
Recall Initiated
October 25, 2019
Terminated
May 20, 2020
Recall Number
Z-0672-2020
Quantity
1 unit
Firm Location
Billerica, MA

Reason for Recall

Medial and lateral polyethylene inserts did not securely fit into the cobalt chrome tray implant.

Distribution

US Nationwide distribution in the state of GA.

Lot / Code Info

Serial Number: 0450585

Root Cause

Under Investigation by firm

Action Taken

Conformis contacted surgeon via phone on 10/16/19 with the replacement details. Replacement inserts that match the implanted tibial tray interlock geometry were shipped to the affected customer.