Identity Imprint PS Tibial Tray Size 4: Lot 540287

Company
Conformis, Inc.
Recall Initiated
March 11, 2024
Posted
April 17, 2024
Recall Number
Z-1591-2024
Quantity
3 units
Firm Location
Billerica, MA

Reason for Recall

Size 5 Tibial Tray from Lot 508287 was packaged in a Size 4 box from Lot 540287. It was confirmed that Lot 540287 and Lot 508287 were swapped between the pouching and final pack.

Distribution

US Nationwide distribution in the states of FL, KS and NH.

Lot / Code Info

Catalog Number: TPS-301-T04S; UDI#: M572TPS301T04S011, Lot number 540287

Root Cause

Process design

Action Taken

On March 22, 2024 customers were sent URGENT MEDICAL DEVICE FIELD ACTION letters via email. Customers were provided the following instructions: Conformis is seeking the return of the devices in boxes marked with Lot 508287 or Lot 540827. This is the only lot that was distributed. Our records indicate you are the recipient of one of the affected Lot. If you have not already, please return the device using the provided FEDEX label. Additionally, return page 2 of this letter to acknowledge receipt of this notification. Please read, complete, and sign page 2 of this letter. If you have any questions please contact myself or another Conformis representative. If you have an urgent question, please call customer service 781 345 9001.