RecallDepth

smith&nephew SMF(TM) HIGH OFFSET STIKTITE(TM) COATED STEM 12/14 TAPER -1H SIZE, STEM, REF 713552511

This recall has been terminated (originally issued February 19, 2019).

Company
Smith & Nephew Inc
Recall Initiated
February 19, 2019
Terminated
May 21, 2021
Recall Number
Z-1029-2019
Quantity
9 units
Firm Location
Memphis, TN
Official Source
View on FDA website ↗

Reason for Recall

Complaints were received indicating that the outer label incorrectly identified the product as size "1" instead of "-1". The product contained within the package is size "-1"

Distribution

US: CA, TN, NC, AR, IN, FL, MO, and, Canada, China, France, Germany, Great Britain, India, Italy, Japan, Mexico, Poland, South Korea, Spain

Lot / Code Info

11CM01091T, 11EM12764T, 11EM12764U, 13MM16036B, 14GM13037D, 15MM09360A, 15MM10833A, 18CM24036 & 18EM15257

Root Cause

Packaging process control

Action Taken

The firm notified its direct accounts by email and letter on 02/18/2019. The notice requested return of the recalled products. Distributors were directed to notify their customers of the recall and ensure that the actions are carried out.