Refinity Short Tip Rotational Intravascular Ultrasound (IVUS) Catheter; Product number: 400-0200.273; Catalog number 89900. Cardiovascular: The REFINITY (or REFINITY ST) Rotational IVUS Catheter (REFINITY or REFINITY ST) is intended for the intravascular ultrasound examination of coronary arteries. Intravascular ultrasound imaging is indicated in patients who are candidates for transluminal interventional procedures. REFINITY (or REFINITY ST) catheters are designed for use as an adjunct to conventional angiographic procedures to provide an image of the vessel lumen and wall structures.

Company
Volcano Corporation
Recall Initiated
March 29, 2017
Posted
April 10, 2017
Terminated
June 21, 2017
Recall Number
Z-1776-2017
Quantity
718 total devices
Firm Location
Rancho Cordova, CA

Reason for Recall

High complaint rates of kinking near the junction of the proximal shaft to telescope section of the catheter. Kinking could result in disruption of normal clinical workflow or loss of imaging.

Distribution

Distribution to US and Japan

Lot / Code Info

Lot # 50069132, Expiry date 6/30/2018; Lot # 50083527, Expiry date 11/30/2018; Lot # 50083533, Expiry Date 11/30/2018.

Root Cause

Under Investigation by firm

Action Taken

Recall letters dated March 28, 2017 will be hand delivered by Volcano reps. Reps will also pick up any unused catheters for return and obtain a signed proof of delivery for both customer and Volcano Rep.