MEVION S250 Radiation therapy system

Company
Mevion Medical Systems, Inc.
Recall Initiated
November 18, 2015
Posted
February 8, 2016
Terminated
April 3, 2017
Recall Number
Z-0767-2016
Quantity
4
Firm Location
Littleton, MA

Reason for Recall

The treatment table positioning belt can slip gear positions, resulting in deviation of patient setup and treatment positions from the time of slippage until the discovery of slippage and the halting of treatment.

Distribution

US nationwide distribution including FL, NJ, OK, and MO.

Lot / Code Info

Manufacturing Lot or Serial Number: S250-0004

Root Cause

Labeling design

Action Taken

Customer notification letter was sent via certified mail on November 18, 2015. The notification explains the issue and actions to be taken by the customer.