Level 1 Normothermic I.V. Fluid Administration Set REF D-60HL

This recall has been terminated (originally issued December 6, 2006).

Company
Smiths Medical Asd, Inc.
Recall Initiated
November 7, 2006
Posted
December 6, 2006
Terminated
November 30, 2009
Recall Number
Z-0250-2007
Quantity
1480 units
Firm Location
Rockland, MA
Official Source
View on FDA website ↗

Reason for Recall

IV Fluid Administration Set may have a blockage of the recirculating warming fluid channel within the disposable administration set, causing an alarm condition of over-temperature during priming.

Distribution

Worldwide, including USA, Japan, Canada, Ireland, UK, Finland, Netherlands, Germany, Austria, Portugal, France, Sweden, Australia, New Zealand, Russia, Italy, Singapore, Kuwait, Saudi Arabia, Czech Republic, Dubai, Spain, Norway, Denmark, and East Temor.

Lot / Code Info

Lot numbers: 1009446, 1035977, 1043174, 1046075, 1048466, 1060983

Root Cause

Equipment maintenance

Action Taken

Smiths Medical ASD notified Direct customers and distributors via letter sent by UPS on or about November 6, 2006 and the days following as necessary until all consignees are notified. Letters to direct customers will be addressed to the Risk Manager. All Smiths Medical affiliates contacted by e-mail and given instructions to send a letter to their customers and to confirm to Smiths Medical that their customers have taken appropriate action by returning the affected product. US Distributors instructed to return the affected product and to provide a customer list. Foreign distributors instructed to return the affected product and given instructions to send a letter to their customers and to confirm to Smiths Medical that their customers have taken appropriate action by returning the affected product.