RecallDepth

SW-2008 Snuggle Warm Sterile Cardiac Connective Warming Blanket; used with the Equator Convective Warmer, are intended for thermal regulation of a patient's temperture and are designed to prevent hypothermia and/ or reduce cold discomfort during and after surgical procedures.

This recall has been terminated (originally issued May 22, 2014).

Company
Smiths Medical Asd, Inc.
Recall Initiated
April 22, 2014
Posted
May 22, 2014
Terminated
February 27, 2015
Recall Number
Z-1650-2014
Quantity
2120 devices
Firm Location
Rockland, MA
Official Source
View on FDA website ↗

Reason for Recall

The hose could separate from the blanket.

Distribution

Worldwide Distribution Including the States of CA, IL, KY, MS, NC, TN, TX, WA, WI and the Countries of BH, DE, NO, RU

Lot / Code Info

Lot: 1006733, 2474599, 2488138, 2523531, 2536250, 2558180, 2563033, 2564447, 2564448, 2615712

Root Cause

Process change control

Action Taken

On April 24, 2014, consignees were notified of the recall via letter. They were instructed to do the following: Inspect inventory and segregate any unused Blankets listed in the notification, complete and return the attached Confirmation Form by fax or e-mail within 7 days of receipt of this notice, and destroy the product.

More recalls by Smiths Medical Asd, Inc.

View all recalls by this company →