Stryker Visum 300 Ceiling-mounted Exam Light.

This recall has been terminated (originally issued August 21, 2013).

Company
Stryker Communications
Recall Initiated
July 25, 2013
Posted
August 21, 2013
Terminated
April 3, 2015
Recall Number
Z-1992-2013
Quantity
1,375 units
Firm Location
Flower Mound, TX
Official Source
View on FDA website ↗

Reason for Recall

Potential incomplete final weld on the drop tube flange.

Distribution

Worldwide Distribution: US natiowide and to the country of Canada.

Lot / Code Info

Product sold under the following part numbers: 0682001410A, 0682001410B, 0682001410C, 0682001410D, 0682001410E, 0682001410F, 0682001410G, and 0682001410H and shipped prior to February 28, 2013.

Root Cause

Process control

Action Taken

The firm decided to recall and notified consignees with an "Urgent Medical Device Correction/Removal" notification letter dated 07/29/13. The product issue was described and recommended actions were provided. Customers may contact Stryker Technical Support at 866-841-5663.