Lenstar LS 900 EyeSuite Biometry version i2.000, Four (4) Protective Cardboard Containers Per Each System. Haag-Streit USA, Inc. A non-invasive, non-contact OLCR (Optical Low Coherence Reflectometer) Biometer used for obtaining ocular measurements and performing calculations to assist in determination of the appropriate power and type of IOL (Intraocular Lens) for implantation after removal of the natural crystalline lens following cataract removal.
- Company
- Haag-Streit Usa Inc
- Recall Initiated
- June 20, 2011
- Posted
- August 9, 2011
- Terminated
- March 8, 2013
- Recall Number
- Z-2964-2011
- Quantity
- 216
- Firm Location
- Mason, OH
Reason for Recall
A patient record may be linked to an incorrect/inconsistent set of calculation results under specific circumstances when using EyeSuite Software version i2.000.
Distribution
Nationwide (USA) Distribution.
Lot / Code Info
Part Number 7220162. Serial Numbers: 21, 127, 223, 224, 284, 285, 316, 318, 319, 320, 322, 323, 324, 325, 326, 327, 328, 330, 331, 332, 333, 364, 365, 366, 368, 369, 371, 372, 376, 378, 379, 381, 384, 385, 386, 405, 407, 408, 409, 411, 412, 413, 414, 415, 416, 417, 418, 419, 420, 421, 423, 424, 443, 446, 447, 448, 449, 450, 452, 453, 454, 455, 456, 457, 458, 459, 460, 463, 469, 470, 481, 482, 483, 484, 485, 486, 487, 489, 490, 491, 492, 493, 494, 495, 496, 497, 499, 500, 527, 528, 529, 531, 532, 533, 535, 536, 537, 538, 539, 540, 541, 542, 543, 544, 545, 546, 547, 548, 549, 550, 551, 552, 558, 559, 560, 561, 563, 564, 565, 566, 567, 568, 569, 570, 571, 572, 573, 574, 576, 578, 579 581, 582, 583, 584, 585, 587, 588, 589, 590, 618, 619, 620, 621, 622, 624, 625, 626, 628, 630, 631, 632, 633, 634, 635, 636, 638, 639, 640, 641, 643, 644, 645, 647, 648, 649, 650, 651, 652, 653, 654, 656, 657, 658, 659, 660, 661, 662, 663, 665, 667, 668, 669, 670, 671, 672, 674, 675, 676, 677, 721, 723, 724, 725, 726, 727, 728, 729, 735, 824, 825, 922, 946, 966, 967, 968, 970, 1018, 1019, 1021, 1022, 1024, 1025, 1114, 1143 & 1195.
Root Cause
Software change control
Action Taken
Haag-Streit USA, Inc. sent an "Urgent Field Safety Notices" dated June 20, 2011 to all affected customers. The letter included a description of the affected product and problem. The firm instructed their customers to make sure all new and current templates are saved and upgrade to EyeSuite version i4.000. Customers are asked to complete and return the Software Update Verification Form. For additional information contact the Haag-Streit at (513) 398-3937.