RecallDepth

Eyesuite 9.3.1 software, LENSTAR, MODEL LS900: ac-powered slit-lamp biomicroscope.

Company
Haag-Streit Usa Inc
Recall Initiated
April 15, 2021
Recall Number
Z-1560-2021
Firm Location
Mason, OH

Reason for Recall

Examination data and patient name may be mixed up when printing or generating a pdf with the interface at Pacific Coast and Laser Institute (PCLI).

Distribution

US Nationwide distribution in the state of WA.

Lot / Code Info

Software version 9.3.1

Root Cause

Software design

Action Taken

On 04/06/2021, Haag-Streit issued an Urgent Medical Device Correction notice to their customer via letter notifying them Eyesuite 9.3.1 software may mix up examination data and patient name when printing or generating a PDF.

More recalls by Haag-Streit Usa Inc

View all recalls by this company →