RecallDepth

Edwards Lifesciences Fem-Flex Femoral Arterial Cannula with Duraflo Coating, DIITFA02225, Sterile, RX only For Single Use Only, Product Usage: Edwards Lifesciences femoral access cannulae are intended for use in situations which require rapid femoral venous and arterial access for short-term (< 6 hours) cardiopulmonary bypass. Vessel access (venous or arterial) is left to the discretion of the physician. Femoral access cannulae may be used in pediatric populations or adult populations based on flow rate requirements and individual patient anatomy. Please consult labeling to determine pressure drop related to flow rates. Extracorporeal circuit components with a Duraflo coating are intended for use in cardiopulmonary surgery when a heparin coated blood path is desired.

This recall has been terminated (originally issued November 1, 2013).

Company
Edwards Lifesciences, Llc
Recall Initiated
August 26, 2013
Posted
November 1, 2013
Terminated
February 22, 2017
Recall Number
Z-0134-2014
Quantity
80 units
Firm Location
Draper, UT
Official Source
View on FDA website ↗

Reason for Recall

Edwards Lifesciences is recalling two lots of Duraflo Cannulas due to reported complaints involving cracks in the t-connector of the devices which were confirmed during product evaluation.

Distribution

Nationwide Distribution.

Lot / Code Info

Lot number 59388992 expiry 1/1/16

Root Cause

Process control

Action Taken

Edwards sent an Urgent Field Safety Notice Product Recall letter dated August 26, 2013 to all affected customer, via Fed-Ex. The product is sold directly to hospitals and distributors. The distributors involved will communicate with their customers directly. The letter identified the affected product, problem and actions to be taken. Customers were instructed to review their entire inventory, quarantine and return the affected products. An acknowledgement form is included for completion to assist in the assessment of the inventory. For questions contact Edwards Customer Service at 800-424-3278.

More recalls by Edwards Lifesciences, Llc