RecallDepth

ThruPort Knot Pusher. This device is sold individually (Model KP1) and as part of the ThruPort Valve Placement Pack (Model VPP). packaged sterile and non-pyrogenic in a sealed, peel-type pouch. The Knot Pusher is a sterile disposable plastic rod used to advance extracorporeal suture knots during minimally invasive cardiac surgery.

This recall has been terminated (originally issued September 18, 2015).

Company
Edwards Lifesciences, Llc
Recall Initiated
August 26, 2015
Posted
September 18, 2015
Terminated
February 10, 2017
Recall Number
Z-2780-2015
Quantity
759 units
Firm Location
Draper, UT
Official Source
View on FDA website ↗

Reason for Recall

The configuration of the slot at the tip of the knot pusher may inhibit intended ease of use. Fraying or splitting of the suture could occur, which could compromise the integrity of the suture knot.

Distribution

Worldwide Distribution - US (nationwide) Distribution and to the countries of : Canada, Japan and Europe.

Lot / Code Info

Model: KP1 - Lot No: 59964698, 59967596, 59984375, 60016312, 60045801, 60045802, 60059982, 60066591, 60089865  Model VPP Lot No: 60022521, 60022522, 60064112

Root Cause

Process control

Action Taken

Edwards sent an Urgent-Product Recall - Action Required letter dated August 26, 2015, to all affected customers. Letters were sent via Fed-Ex. Please check your inventory for the affected Lot Numbers listed above and return the enclosed confirmation form. If your inventory includes affected product, please contact Edwards Lifesciences to obtain an RGA number and return the affected product. We apologize for the inconvenience caused by this action and appreciate your attention to this matter. If you have questions that have not been answered by this letter, please call Edwards Customer Service at (800) 424-3278 from 6:00AM 4:30PM Pacific Time.