Acuson X300 systems at software versions 1.00.00 to 1.0.10 and 2.0.00 to 2.0.02, model numbers 10037409, 10132987, and 10038837, manufactured by Siemens Medical Solutions ultrasound business unit, Mountain View, CA Medical device is an ultrasound imaging system for general radiology, fetal, abdominal, intraoperative, pediatric, small parts, neonatal/adult cephalic, cardiac, transesophageal, pelvic, transcranial, ob/gyn, urology, vascular, musculoskeletal, superficial musculoskeletal and peripheral vascular applications. The system also provides for measurements of anatomical structures and for analysis packages that provide information that is used for clinical diagnosis purposes.

This recall has been terminated (originally issued April 25, 2011).

Company
Siemens Medical Solutions Usa, Inc.
Recall Initiated
July 30, 2009
Posted
April 25, 2011
Terminated
April 25, 2011
Recall Number
Z-2065-2011
Quantity
2312 units
Firm Location
Mountain View, CA
Official Source
View on FDA website ↗

Reason for Recall

Software Error with potential for misinterpretation of data-- The previous patient's last capture clip image can appear in the next patient's study.

Distribution

Worldwide distribution: USA (nationwide) and countries including: Albania, algeria, Argentina, Australia, Austria, Azerbaijan, Belgium, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile,China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Ecuador, Egypt, El Salvador, Estonia, Finland, France, Georgia, Germany, Greece, Hungary, India, Indonesia, Iran, Ireland, Italy, Japan, Kazakhstan, Kenya, Kosovo, Lativia, Lithuania, Luxembourg, New, Zealand, Nicaragua, Norway, Paraguay, Peru, Philippines, Poland, Portugal, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Venezuela, Viet Nam.

Lot / Code Info

Serial numbers

Root Cause

Software design

Action Taken

The firm, Siemens, sent a "CUSTOMER SAFETY ADVISORY NOTIFICATION" letter to its customers on July 30, 2009. The letter described the product, problem and actions to be taken. The customers were instructed they need to update any of the patient registration settings to reset the cine buffer so that no data is transferred between studies, if the system software is not being updated. Siemens informed the customers that a mandatory update software is being installed on their system free of charge. If you have any questions, contact (650) 969-9112.