Stryker Equipment Delivery System (EDS); Ceiling Mounted Equipment Booms; Catalog Number: 0682000383, 0682000384, 0682000385(A-D), 0682000386(A-D), 0682000426, 0682000429(A-D), 0682000326, 0682000311, 0682000361, 0682000362, 0682000230, 0682000337, 0682000338, 0682000232, 0682000233, 0682000339, 0682000234, 0682000235, 0682000236, 0682000340, 0682000341, 0682000238, 0682000239, 0682000342, 0682000240, 0682000241, 0682000931, 0682000343, 0682000344, 0682000328, 0682000933, 0682000345, 0682000329, 0682000330, 0682000934, 0682000346, 0682000347, 0682000331, 0682000936, 0682000348, 0682000332, 0682000333, 0682000401, 0682000302, & 0682000327. The booms are ceiling mounted devices. They are available in articulating and non-articulating configurations, in single or tandem (pair) configuration, with different shelves configuration to allow for various load capacities. Intended for use as ceiling-mounted devices for supporting or positioning equipment in the surgical suite and/or delivers gases or electricity to these devices.

This recall has been terminated (originally issued May 21, 2010).

Company
Stryker Communications Corp
Recall Initiated
April 15, 2010
Posted
May 21, 2010
Terminated
January 24, 2011
Recall Number
Z-1641-2010
Quantity
608 (booms and lights)
Firm Location
Flower Mound, TX
Official Source
View on FDA website ↗

Reason for Recall

Installation records are incomplete.

Distribution

Nationwide Distribution

Lot / Code Info

Lot Code: Product Installed Between January 1, 2008 and August 31, 2009.

Root Cause

Employee error

Action Taken

An Urgent: Medical Device Product Correction customer notification letter, dated 4/15/2010, has been sent out to all affected accounts via regular mail in the form of a certified, return request receipt letter. Accounts will receive a follow-up phone call from a Stryker Representative to schedule a re-inspection of equipment to see if it mets established critical specifications or not. The firm recommends that customers continue using all Stryker Communications products listed in the notice. Questions can be directed to Katie Blackwell at 972-410-7660 or via email at StrykerCommunicationsRA@stryker.com.

More recalls by Stryker Communications Corp

View all recalls by this company →