RecallDepth

Exacta-Mix 2400, OS v1.07, Model No. 8300-0073, Pharmacy Compounding System, Baxa Corporation, Englewood, CO 80112.

This recall has been terminated (originally issued September 27, 2007).

Company
Baxa Corporation
Recall Initiated
June 27, 2007
Posted
September 27, 2007
Terminated
December 7, 2007
Recall Number
Z-1238-2007
Quantity
30 units
Firm Location
Englewood, CO
Official Source
View on FDA website ↗

Reason for Recall

A software defect could allow up to 50mL of extra volume being added to a TPN solution.

Distribution

Class I Recall-Worldwide Distribution: USA including states of CA, CO, FL, KY, MI, NY, OH, OR, and TX and the UK

Lot / Code Info

Version 1.07 of the EM2400 Operating Software. (Serial Numbers: 36805,36806,36356,36399,36804,36799,36797,36398,32438,36798,36460,36794,32346,36800,36801,36802,36803,(37767-36773), (36788-36793),36795 and 36796 (foreign)

Root Cause

Other

Action Taken

All customers were notified via telephone on 06/27/2007 with follow-up of a faxed/emailed Urgent Product Recall letter dated 27 June 2007, detailing the software defect. End users were given a work around procedure to use until conversion back to version 1.06 could be scheduled. An additional copy of the recall letter was sent by FedEx.