AMPLATZER Duct Occluder, Order No. 9-PDA-004, Device Size 6/4mm, Length 7 mm. Sterile EO. Single Use Only. Manufactured by AGA Medical Corporation, 682 Mendelssohn Avenue, Golden Valley, MN 55427 USA. The AMPLATZER Duct Occluder is a self-expandable device made from a Nitinol wire mesh. A retention skirt on the aortic side provides secure positioning in the ampulla of the ductus. As the occluder is implanted, it expands outward and the wires push against the wall of the ductus. Polyester fabric is sewn into the occluder with polyester thread. The fabric induces thrombosis that closes the communication.

This recall has been terminated (originally issued January 31, 2006).

Company
Aga Medical Corporation
Recall Initiated
September 6, 2005
Posted
January 31, 2006
Terminated
December 26, 2006
Recall Number
Z-0462-06
Quantity
18,725 total
Firm Location
Golden Valley, MN
Official Source
View on FDA website ↗

Reason for Recall

The pointed end of a screw securing the occluder to the delivery cable could scrape off microscopic fragments of the polytetrafluorethylene (PTFE) plastic lining of the sheath. Those shavings could potentially migrate into a patient's bloodstream.

Distribution

OUS to the following countries: Algeria, Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Rica, Denmark, El Salvador, Finland, France, Germany, Greece, Guatemala, Hungary, India, Indonesia, Isreal, Italy, Jordan, Korea, Kuwait, Malaysia, Mexico, Morocco, Netherlands, Norway, Pakistan, Peru, Philippines, Poland, Portugal, Saudia Arabia, Singapore, Slovakia, South Africa, Spain, Sweeden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Uruguay, Venezuela, Viet Nam and United Kingdom.

Lot / Code Info

all lots

Root Cause

Other

Action Taken

User facilities and distributors were notified via email beginning 09/06/05. Additionally a hard copy of the recall letter was sent via certified mail or Federal Express. AGA requests that the Response Form be completed and faxed or emailed back to them. AGA Medical staff will assist users in returning product distributed within the US.