AGA Medical Corporation AMPLATZER Delivery System, 10 French, 45 Degree Angle Curve, 80 Usable Length (cm), REF 9-DEL-10F-45/80. Sterile EO. AGA Medical Corporation, 682 Mendelssohn Avenue, Golden Valley, MN 55427 USA. Manufactured in U.S.A. The AMPLATZER Delivery System was designed to facilitate attachment, loading, delivery and deployment of the AMPLATZER Septal Occluder and comprised of a delivery sheath, dilator, loading device, plastic vise and delivery cable.

This recall has been terminated (originally issued September 3, 2008).

Company
Aga Medical Corporation
Recall Initiated
April 25, 2008
Posted
September 3, 2008
Terminated
October 13, 2010
Recall Number
Z-2192-2008
Quantity
10 (9 US, 1 OUS)
Firm Location
Golden Valley, MN
Official Source
View on FDA website ↗

Reason for Recall

AGA Medical Corporation has identified an issue with outer pouch seal integrity within one production lot of 10 French 45/80 delivery systems and one lot of 24 mm Sizing Balloon. We have not observed incomplete seals with the inner pouch. An intact inner pouch seal will maintain device sterility. AGA is asking that customers inspect the specific lot of devices to confirm the integrity of the ou

Distribution

Within the US to include: AL, AZ, CA, CT, FL, IA, ID, MA, MD,MI, MO, NC, NH, NY, OH, PA, SD, TN, TX, WI. OUS to include: Brazil, Greece, Netherlands, Portugal, and Slovakia

Lot / Code Info

lot number M07K29-32

Root Cause

Packaging process control

Action Taken

The firm Fed Ex'd and emailed a Field Inspection/Correction letter to customer 04/28/2008. The letter identified affected product, requested that customers check the outer package seal, and if it is incomplete the device should be returned for replacement. A form was also asked to be returned to AGA Medical. Contact AGA Medical at 1-763-513-9227 for assistance.