OPTIMA FW (polymacon) Visibility Tinted Contact Lenses, labeled as -2.25D, 8.7 mm. Responsible firm on the label: Bausch & Lomb Inc., Rochester, NY 14692. Each box contains six lenses.
This recall has been terminated (originally issued December 19, 2002).
- Company
- Bausch & Lomb Inc
- Recall Initiated
- November 26, 2002
- Posted
- December 19, 2002
- Terminated
- June 23, 2003
- Recall Number
- Z-0348-03
- Quantity
- 2,081 / 6-packs
- Firm Location
- Rochester, NY
- Official Source
- View on FDA website ↗
Reason for Recall
Lenses labeled as -2.25 diopter contain -0.75 diopter lenses.
Distribution
Nationwide distribution.
Lot / Code Info
Lot R21000211
Root Cause
Other
Action Taken
Letters dated 12/2/02 issued to distributors and practitioners requesting return of the affected lot.