Surgical Table Headrest, Catalog/Part #0788-300-012, component of VERTIER Surgical Table. Distributed as part of the VERTIER Surgical Tables or individually by Stryker Communications, Flower Mound, TX 75028. The Vertier surgical table is intended for use in general surgery or diagnostic examinations to position a patient. The Vertier Surgical Table is an AC powered, general purpose, mobile surgical table with electro-hydraulically controlled movements which are part of a control feedback system. The headrest is included with each table.

This recall has been terminated (originally issued May 26, 2009).

Company
Stryker Communications Corp
Recall Initiated
May 8, 2009
Posted
May 26, 2009
Terminated
April 26, 2010
Recall Number
Z-1418-2009
Quantity
230 headrests.
Firm Location
Flower Mound, TX
Official Source
View on FDA website ↗

Reason for Recall

Potential for health care provider/hospital staff member to sustain injury to fingers or hands when removing/handling the headrest component of the VERTIER Surgical Table.

Distribution

Nationwide Distribution and Canada.

Lot / Code Info

All surgical tables with headrest, Catalog/Part #0788-300-012. No Lot numbers.

Root Cause

Device Design

Action Taken

An "URGENT: Device Recall" letter dated May 11, 2009 was issued to consignees. Consignees were informed that Stryker service representatives will visit each location to affix a caution label identifying the pinch point on each headrest. Additionally, consignees will be provided with an updated surgical table user manual. Direct questions to Stryker Communications Corporation by calling 972-410-7100.