Operon Electrical Operating Room Table. The device is used as an electrical operating room table.

This recall has been terminated (originally issued April 27, 2009).

Company
Berchtold Corp.
Recall Initiated
February 2, 2009
Posted
April 27, 2009
Terminated
June 4, 2009
Recall Number
Z-1123-2009
Quantity
453 units
Firm Location
Charleston, SC
Official Source
View on FDA website ↗

Reason for Recall

Difficulty in pressing the button needed to release a bed part resulted in an injury to a caregiver's finger which became stuck in metal latch under the gas cylinder of the bed piece.

Distribution

Worldwide Distribution -- US and Canada.

Lot / Code Info

D 850, D750, D 752 and B 810.

Root Cause

Device Design

Action Taken

An official notification letter dated February 3, 2009 was hand-carried by a local Berchtold Corporation Technical Service representative to each account. The letter indicated that the Berchtold Corporation Service Technician will complete recommended upgrades. Direct questions concerning this recall to your BERCHTOLD Technician Service at 1-800-243-5135 and press the Number 2.