Hand Controls for Operon D750, D752, D760, D820 and D850 The hand controls are used to operate OPERON surgical tables

This recall has been terminated (originally issued June 6, 2013).

Company
Berchtold Corp.
Recall Initiated
March 12, 2013
Posted
June 6, 2013
Terminated
May 8, 2017
Recall Number
Z-1488-2013
Quantity
3763
Firm Location
Charleston, SC
Official Source
View on FDA website ↗

Reason for Recall

The firm received complaints for unintended movements.

Distribution

Nationwide Distribution

Lot / Code Info

The hand controls are neither serialized nor lot coded.

Root Cause

Device Design

Action Taken

Berchtold sent an Urgent Medical Device Recall letter in March 2013, to all customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were recommended to clean the affected product and use the instructions to reduce the possibility of unintended movements until the firm could replace the affected product. Customers with questions were advised to contact the Berchtold Operon Pendanct Recall Coordinator at 800-243-5135, or 843-569-6100, ext 306. For questions regarding this recall call 843-569-6133.