Excelart (tm) VANTAGE P2 & P3, Magnetic Resonance Imaging System,

This recall has been terminated (originally issued November 2, 2004).

Company
Toshiba American Med Sys Inc
Recall Initiated
September 16, 2004
Posted
November 2, 2004
Terminated
July 13, 2005
Recall Number
Z-0061-05
Quantity
22
Firm Location
Tustin, CA
Official Source
View on FDA website ↗

Reason for Recall

Scanned images are acquired at a position that is shifted relative to the intended position in the slice direction due to a software anomaly.

Distribution

PA, MD, FL, OH, NY, CA, AR, KY, OK, AZ, TX, NV, IN.

Lot / Code Info

Serial numbers: A4512006, A452017, A4532011, A3612004, A4532014, A4572020, A4572021, A3622005, A4562003, A3592003, A4562018, A4522008, A4532010, A4542015, A4532013, A3622001, A3582001, A4522009, A4532012, A3592002, A4522007, A4552016.

Root Cause

Other

Action Taken

Firm issued a letter to users on 9/29/2004 which includes a method to avoid the software problem until it can be corrected by an update.