Microstream SMART CAPNOLINE PLUS O2 100UN Part Number: 010210 Oridion 3012176-112 Physio Control

This recall has been terminated (originally issued July 13, 2010).

Company
Oridion Medical
Recall Initiated
May 26, 2010
Posted
July 13, 2010
Terminated
June 21, 2012
Recall Number
Z-2007-2010
Quantity
1119 boxes
Official Source
View on FDA website ↗

Reason for Recall

The Microstream Filterline may not be recognized by the monitor when connected. No CO2 readings will be displayed and no alarms may sound for unmonitored patients. Replacing the filterline may delay treatment.

Distribution

AZ, CA, FL, HI, LA, , MA, MO, NJ, OH, PA, WA, WI,

Lot / Code Info

Lot numbers from Q0901006 to Q0908223. Produced between January and July 2009,

Root Cause

Process design

Action Taken

Oridion notified customers by letter on 5/26/10 and requested that they remove the FilterLines from the related production lots from the field and an immediate replacement with FilterLines from subsequent production lots. Questions are directed to the responsible person at the supplier named in the letter, or to Oridion Capnography at 888-674-3466.