Micromix BCR Compounder, product code 2M8297; a pharmacy automation device with a barcode reader used in preparation of parenteral solutions; Baxter Healthcare Corporation, Clintec Nutriiton Division, Deerfield, IL 60015 USA
This recall has been terminated (originally issued January 29, 2004).
- Company
- Baxter Healthcare Corp. Rt
- Recall Initiated
- December 16, 2003
- Posted
- January 29, 2004
- Terminated
- April 18, 2005
- Recall Number
- Z-0377-04
- Quantity
- 6 units
- Firm Location
- Round Lake, IL
- Official Source
- View on FDA website ↗
Reason for Recall
The compounder may omit an ingredient during the compounding process.
Distribution
Wisconsin and Utah.
Lot / Code Info
Serial numbers T2001/T20014, T20016/T20018, T20021, T20023/T20028
Root Cause
Other
Action Taken
Baxter telephoned the accounts on 12/16/03, informing them of the potential for the Micromix BCR to omit an ingredient during the compounding process. The accounts were instructed to cease use of the Micromix BCR and return the units to Baxter via next day delivery.