i-view video laryngoscope. Model Number: 8008000. The i-view video laryngoscope is indicated to facilitate laryngoscopy.

Company
Intersurgical Inc
Recall Initiated
June 17, 2025
Posted
July 11, 2025
Recall Number
Z-2102-2025
Quantity
2,152 pieces
Firm Location
East Syracuse, NY

Reason for Recall

Potential for faulty devices as a result of depleted batteries.

Distribution

US distribution AZ, CA, FL, GA, IL, IN, KS, MD, MI, MO, NV, NY, OH, PA, SC, TX, VA.

Lot / Code Info

Model Number: 8008000. UDI-DI (Case): 05030267166497. UDI-DI (Single Unit): 5030267150649. Lot Numbers: 1240555, 1240793.

Root Cause

Component design/selection

Action Taken

Intersurgical notified consignees and account executives via email on 06/17/2025. The notification instructed consignees and account executives to immediately discontinue use and quarantine any affected stock on hand, notify customers if affected units have been further distributed, and contact Intersurgical Customer Support at 800-828-9633 for a Return Goods Authorization (RGA) Number and return label to return all affected units. Also, they were instructed to complete and return the Customer Product Recall Acknowledgement Form.