RecallDepth

One-piece Guedel airway, size 3, ISO 9.0, yellow. Model Number: 1113090.

Company
Intersurgical Inc
Recall Initiated
June 20, 2025
Posted
August 6, 2025
Recall Number
Z-2276-2025
Quantity
8,640 pieces
Firm Location
East Syracuse, NY

Reason for Recall

Potential contamination with small burrs, which if detach could be inhaled and result in potential complications such as airway obstruction, tissue irritation, inflammation and infection.

Distribution

US distribution to AZ, CA, CT, FL, FL, HI, MA, MI, OH, UT, VA.

Lot / Code Info

Model Number: 1113090. UDI-DI (Case): 05030267153961, UDI-DI (Single Unit): 5030267050680. Lot Number: 32420657

Root Cause

Process control

Action Taken

Intersurgical notified consignees on 06/20/2025 via email with recall letter. Consignees were instructed to immediately discontinue use and quarantine any affected units, immediately notify customers if product has been further distributed, arrange for the return of affected units and complete the Customer Product Recall Acknowledgement Form.