RecallDepth

Medline procedure kits labeled as: 1) Port Access Kit, Reorder number CVI5130; 2) VANTEX 7FR 3L 20CM CVC STOCK+ BUNDLE US. Reorder number STCVC01USGV; general surgical tray

This recall is currently active, issued January 28, 2025. It was issued by Medline Industries, Lp - Northfield.

Company
Medline Industries, Lp - Northfield
Recall Initiated
November 18, 2024
Posted
January 28, 2025
Recall Number
Z-1040-2025
Quantity
749 units
Firm Location
Northfield, IL
Official Source
View on FDA website ↗

Reason for Recall

Certain Medline kits were manufactured and distributed with Procedure Masks with SO SOFT Earloops, and SOFT TOUCH II Surgical Masks which were subsequently recalled for not meeting the bioburden testing requirements of EN 14683.

Distribution

US Nationwide distribution.

Lot / Code Info

1) CVI5130: UDI/DI 10653160372349 (each) 00653160372342 (case), Lot Numbers: 2024040490, 2024073190, 2024082790, 2024101790,; 2) STCVC01USGV: UDI/DI 10653160376965 (each) 00653160376968 (case), Lot Numbers: 2024050990.

Root Cause

Under Investigation by firm

Action Taken

Medline issued a MEDICAL DEVICE RECALL notice to its consignees on 11/18/2024 via USPS and email. The notice explained the reason for the recall and requested quarantine of affected products pending correction. The firm requested the completion of the response form, which upon receipt, Medline will send labels to affix to the affected products with the instructions for the consignee to remove the affected component prior to using the kit.