Exactech Alteon Hip XLE Liner, Catalog Numbers: a) 01-030-40-0636, b) 01-030-40-0640, c) 01-030-42-0536

Company
Exactech, Inc.
Recall Initiated
November 21, 2023
Posted
January 18, 2024
Recall Number
Z-0578-2024
Firm Location
Gainesville, FL

Reason for Recall

Exactech has received 3 complaints for 4 devices related to a loss of vacuum in the inner-most vacuum bag.

Distribution

US NY, TX, IN, IL, MN, KY, VA, MD, TN, GA, NH, ME, SC, NV, FL, AR and Australia, Brazil, Chile, France, Germany, Guatemala, Italy, Japan, Korea, Slovenia, Spain, UK

Lot / Code Info

a) 01-030-40-0636, GTIN 10885862549655, Serial Numbers: A678476, A678485, A678486, A678495, A678508, A678516, A678819, A678820, A678822, A678826, A678827, A678832, A678845, A678847, A678855, A678859, A678863; b) 01-030-40-0640, GTIN 10885862549662, Serial Numbers: A746645, A746649; c) 01-030-42-0536, GTIN 10885862549945, Serial Numbers: A655570, A655573, A655574

Root Cause

Packaging change control

Action Taken

Exactech issued an URGENT MEDICAL DEVICE RECALL notice to its direct consignees on 11/21/2023 via email. The notice explained the issue, clinical impact, and requested the removal and return of all affected products. 11/22/2023, the firm initiated the recall OUS also by email.