epoc BGEM BUN Test Card with epoc Host SW v3.37.3, epoc NXS SW v4.10.6, Sensor Configuration 41.1

Company
Siemens Healthcare Diagnostics Inc
Recall Initiated
June 1, 2023
Posted
August 28, 2023
Recall Number
Z-2476-2023
Quantity
3201 ea
Firm Location
Norwood, MA

Reason for Recall

There is potential for discrepant high glucose results in samples with glucose results on the lower end of the reportable range.

Distribution

US distribution to AL, AR, AZ, CA, CO, DC, DE, FL, IA, ID, IL, KS, KY, LA, MI, MN, MO, NC, ND, NH, NY, OH, OK, SC, SD, and TX OUS distribution to Canada, China, Colombia, Finland, India, Mexico, Norway, Sweden, United Arab Emirates, and Vietnam

Lot / Code Info

UDI-DI: 00809708121860 Siemens Material Number: 10736515 Lots: 04-23003-60 (expiry 20-Jun-2023), 11-22337-60 (expiry 20-May-2023), 04-23034-40 (expiry 21-Jul-2023), 04-23059-50 (expiry 15-Aug-2023), 04-23059-60 (expiry 15-Aug-2023), and 04-23069-40 (expiry 25-Aug-2023)

Root Cause

Under Investigation by firm

Action Taken

Consignees were notified via an Urgent Medical Device Correction letter beginning June 5, 2023. Consignees were informed of the issue and were provided a workaround and actions to take. The next sensor configuration/software version, sensor configuration 42.1/epoc Host SW v3.38.2 /epoc NXS SW v4.11.11, is available to customers in Siemens Healthineers Document Library or the epoc Live Update Service (eLUS). This next sensor configuration, 42.1, mitigates the issue.