Visera Hysterovideoscope Olympus HYF Type V

This recall is currently active, issued August 29, 2023. It was issued by Olympus Corporation Of The Americas.

Company
Olympus Corporation Of The Americas
Recall Initiated
July 20, 2023
Posted
August 29, 2023
Recall Number
Z-2480-2023
Quantity
363 units
Firm Location
Center Valley, PA
Official Source
View on FDA website ↗

Reason for Recall

IFU labeling is being updated to revise EtO gas mixture and elimination of reusable cleaning brushes.

Distribution

US Nationwide distribution including in the states of AL, AZ, CA, CO, CT, FL, GA, IA, IL, IN, KS, LA, MA, MD, MI, MN, MO, NC, ND, NE, NJ, NM, NV, NY, OH, OR, PA, RI, SC, SD, TX, UT, WA, and WI.

Lot / Code Info

Model HYF-V, all serial numbers

Root Cause

Reprocessing Controls

Action Taken

A customer letter was issued on July 20, 2023. Customers were advised that the Instructions for use were updated. Customers are asked to complete a response form of acknowledgement and submit to Sedgwick at olympus8606@sedgwick.com or Fax 866-557-1242.