RecallDepth

Medline Pill Splitters; 1 each/box; 144 boxes/carton

Company
Medline Industries, Lp - Northfield
Recall Initiated
May 11, 2022
Recall Number
Z-1287-2022
Quantity
840,087 each
Firm Location
Northfield, IL

Reason for Recall

Product lid (which houses a blade) is coming unhinged from the base, thus exposing the blade.

Distribution

Worldwide distribution - US Nationwide and the countries of Canada, Singapore, Panama, and Hongkong.

Lot / Code Info

Model Number: NON135000; UDI/DI:8019673489; Lot Numbers: 50920060001, 50920080001, 50920110001, 50921030001, 50921040001, 50921050001, 50921060001, 50921080001, 50921090001, 50921100001, 50921110001, 50921120001, 50922020001

Root Cause

Process change control

Action Taken

The firm issued a recall notice to consignees on 05/11/2022 via mail. The letter instructs the consignee to check their stock, destroy affected product, and return the response form. On 5/16/2022, the firm sent out an updated recall notice.