RecallDepth

Medline Contro-Bulb Syringe, REF DYND20125; bulb irrigation syringe

This recall is currently active, issued September 30, 2024. It was issued by Medline Industries, Lp - Northfield.

Company
Medline Industries, Lp - Northfield
Recall Initiated
January 4, 2024
Posted
September 30, 2024
Recall Number
Z-1157-2024
Quantity
4831 units
Firm Location
Northfield, IL
Official Source
View on FDA website ↗

Reason for Recall

Packaging material was updated for the Bulb Irrigation Syringe (DYND20125), and additional testing done at Medline Corporate indicated the potential for the sterile barrier to be breached during transportation. This only affects two lot numbers (96922110001 and 96923100001).

Distribution

US, Nationwide

Lot / Code Info

UDI/DI 40080196031765 (case), 10080196031764 (unit) Lot Numbers: 96922110001 and 96923100001

Root Cause

Under Investigation by firm

Action Taken

Medline Industries, LP issued an IMMEDIATE ACTION REQUIRED notice to its consignees on 01/04/2024 via USPS First Class. The notice explained the problem with the device, risk involved, and requested the consignee quarantine then destroy the affected product. Upon receipt of the completed destruction form, firm will issue credit, if applicable. Consignee was directed to notify all parties of this action to whom the product was distributed. For questions, contact 866-359-1704