RecallDepth

Medline Kits, trays, and packs labeled as follows: a) C L PREP, REF DYNJ69664; b) DMEK PREP PACK, REF DYNJ81659A; c) PRE OP KIT 1, REF DYKS1377; d) PRE OP KIT 1, REF DYKS1377A

This recall is currently active, issued February 25, 2024. It was issued by Medline Industries, Lp - Northfield.

Company
Medline Industries, Lp - Northfield
Recall Initiated
November 15, 2023
Posted
February 25, 2024
Recall Number
Z-1116-2024
Quantity
1270 units
Firm Location
Northfield, IL
Official Source
View on FDA website ↗

Reason for Recall

Medline Industries, LP is recalling certain kits and trays that were manufactured using specific lots of Nurse Assist (as components): 0.9% Sodium Chloride Irrigation USP, Sterile Water for Irrigation USP and Saline Flush Syringe. Nurse Assist issued a recall due to the potential lack of sterility, which could result in the solution being non sterile.

Distribution

Distribution US nationwide including Puerto Rico, Canada, UAE, and Panama

Lot / Code Info

a) REF DYNJ69664, UDI/DI 40193489966337 (case), 10193489966336 (unit), Lot Numbers: 21LBW095, 22EBT126, 22HBQ767, 22LBB063, 23EBF143; b) REF DYNJ81659A, UDI/DI 40195327277094 (case), 10195327277093 (unit), Lot Numbers: 22KBK040; c) REF DYKS1377, UDI/DI 40195327216673 (case), 10195327216672 (unit), Lot Numbers: 22LLA282; d) REF DYKS1377A , UDI/DI 40195327216673 (case), 10195327216672 (unit), Lot Numbers: 23BLA389

Root Cause

Nonconforming Material/Component

Action Taken

MEDLINE issued an IMMEDIATE ACTION REQUIRED NURSE ASSIST PRODUCT RECALL notice to its consignees on 11/16/2023 via USPS first class mail and email. The notice explained the problem, requested log on to their portal to identify the products and check their inventory so the product could be quarantined. Once Medline receives the completed response form, over-labels will be shipped to the consignee. Medline Industries, LP instructed customers to over-label the convenience kits with a warning label indicating to remove the sodium chloride irrigation or Sterile water and replace from a different sterile supply. Distributors or consignees who further distributed the affected products were directed to notify their customers. For questions, contact the Recall Department at 866-359-1704 or recalls@medline.com.