The Vantage Total Ankle Tibial Insert, Fixed Bearing, Size 4, Right 10mm.

This recall is currently active, issued August 10, 2020. It was issued by Exactech, Inc..

Company
Exactech, Inc.
Recall Initiated
August 10, 2020
Recall Number
Z-2926-2020
Quantity
11 units
Firm Location
Gainesville, FL
Official Source
View on FDA website ↗

Reason for Recall

The Vantage Total Ankle Tibial Insert, Fixed Bearing, Size 4, Right 10mm, were found to have a labeling error. Exactech is recalling the Vantage Total Ankle Tibial Insert, Fixed Bearing, Size 4, Right 10mm because they were found to have a labeling error. The upper right corner of the label incorrectly identifies the implant size as 3 when the implant size is 4.

Distribution

Domestic: CO,FL,OK, PA. International: Chile

Lot / Code Info

Catalog Number: 350-22-44 Lot numbers: 05144883 and 05144884 Serial Numbers: 6417984, 6417989, 6417996, 6417997, 6417998, 6418002, 6418003, 6488668, 6488673,6488674,6488675.

Root Cause

Error in labeling

Action Taken

On August 18, 2020 Exactech sent an Urgent Product Recall Notice to all its consignees with the following instructions: -Immediately cease distribution or use of this product. -Extend this information to your accounts that may have this product in their possession. - Identify and quarantine any of the subject devices in your inventory (including loaner kits). -Complete and returned the attached Recall Inventory Response Form to Exactech via email at recalls@exac.com.